Installation and wiring
Type 8045
Correct
flow direction
Incorrect
Correct Incorrect
Fig. 7: Air bubbles within the pipe
If necessary, use a flow conditioner to improve measurement
precision.
7.3. Installation into the pipe of a
8045 with a G2'' nut
In order to ensure a high accuracy of the measurements and
good stability of the „flow zero“ point, install the device into
the processed medium at least 24 H before calibration.
Install the S020 fitting into the pipe taking into account the rec-
ommendations in chap. 7.2.
Check that there is a seal on the fitting and that it is not damaged.
Replace the seal if necessary.
Check that there is a seal (see mark 6, Fig. 8) on the flow sensor.
Insert the nut (see mark 3, Fig. 8) on the fitting.
Insert the snap ring (mark 2 Fig. 8) into the groove (mark 5 Fig.
8).
Position the arrow on the side of the device in the direction of the
flow: the totalizers will increment.
Insert the device (mark 1 Fig. 8) into the fitting.
Tighten the nut (mark 3, Fig. 8) by hand on the device.
English